Hotline: 0988.816.815     vie    eng

Digital refractometers without internal temperature control - DR6000

Digital refractometers without internal temperature control - DR6000
zoom Zoom image
Price: 0 VND
Manufacturer: Kruss - germany
Quantity: 

Description

KEY FEATURES

DR6000 - Specialists for sugar containing solutions

The DR6000 models are especially suited for highly accurate measurements of products from the sugar industry. Instead of temperature control, the temperature compensation according to ICUMSA can be used. The digital refractometer models DR6000 without internal Peltier temperature control are especially suited for highly accurate measurements of products from the sugar industry. Instead of the temperature control, you can use the temperature compensation according to ICUMSA. There are four model variants available for the different requirements on accuracy, resolution and measuring range, whereby the same operation and functionality can be accessed.

  • Measuring all sugar containing samples
  • Built-in thermostat connections for temperature control with external circulating thermostat
  • Preinstalled and freely definable scales with conversions bases on tables or formulas
  • Chemical-resistant materials such as measurement prism made of sapphire, measuring tray made of stainless steel
  • Intuitive operation via state-of-the-art touch-screen display
  • Complete data recording and backup including audit trail
  • Flexible data export
  • Compliance with global standards and norms

APPLICATIONS

The refractometers of the DR6000 models are the ideal solution for the sugar industry

  • Determination of the sugar concentration in semi-finished products and end products
  • Determination of the solids content in solutions
  • Determination of the purity in combination with a polarimeter

 

Conformity with national and international standards

Refractometers are also designed and used in accordance with standards. Standards in refractometry regulate, under the aspect of “best practice”, the environmental conditions and sample preparation as well as measurement tolerances, properties and equipment features of a device or certificates of calibration equipment and testing instruments. Whoever uses refractometers according to normative specifications can guarantee that measurements are correct and work according to reproducible procedures.

 

All samples are measurable

From bright to dark, from basic to acidic, from a low to a high viscosity – for our refractometers almost all samples are measurable. A high-quality optics, intelligent evaluation logarithms and a powerful LED enable the measurement of particularly challenging samples. These include z. For example, non-homogeneous samples such as fruit juices containing pulp, opaque samples such as heavy oils or emulsions that are common in cosmetics. Aggressive acids and bases can be easily measured using a chemical-resistant stainless steel measuring tray and a sapphire prism.

In order to serve the existing variety of samples and the wide range of applications, we offer a wide range of different refractometers. For the selection of the suitable refractometer, it is helpful to define the required measuring range.

In the following table we have compiled an overview of typical samples from our customers and sorted them by refractive index. Most samples were measured at standard measurement conditions (20 ° C, λ = 589 nm), deviations are indicated. The measurements given are not specifications.

 

Model specifications

 

DR6000

DR6100

DR6200

DR6300

MEASUREMENT DATA

 

SCALES

Refractive index (nD)
Concentration of sucrose, glucose, fructose and invert sugar [%Brix],
User-defined

MEASUREMENT RANGE

nD 1.3200–1.5800
0–95 %Brix

nD 1.3200–1.7000
0–95 %Brix

nD 1.32000–1.58000
0–95 %Brix

nD 1.32000–1.70000
0–95 %Brix

MEASUREMENT ACCURACY

nD ±0.0001
±0.1 %Brix

nD ±0.00002
±0.02 %Brix

RESOLUTION

nD 0.0001
0.1 %Brix

nD 0.00001
0.01 %Brix

MEASUREMENT PERIOD

approx. 4 s

MEASUREMENT PRISM

Sapphire

LIGHT SOURCE

LED

WAVELENGTH

589 nm

TEMPERATURE DATA

 

TEMPERATURE CONTROL

Without a sample temperature control, you can connect a thermostat for the temperature control

TEMPERATURE COMPENSATION

Can be activated
(ICUMSA or freely definable)

TEMPERATURE MEASUREMENT

With integrated temperature sensor Pt100

TEMPERATURE MEASUREMENT RANGE

10–80 °C

TEMP. MEASUREMENT ACCURACY

±0.1 °C

TEMP. MEASUREMENT RESOLUTION

0.1 °C

Application


Accessories


Video


Catalogue and User Manual


Download Catalogue
Download User Manual
Copyright MAXXmarketing Webdesigner GmbH
unnamed