ECS 8060 Nitrogen, Protein Analyzer

ECS 8060 Nitrogen, Protein Analyzer
zoom Zoom image
Price: 0 VND
Manufacturer: NC TECH - Italy
Quantity: 

Description

ECS 8060 is a Nitrogen, Protein Analyzer Model making use of the Dumas combustion method. ECS 8060 is a state-of-the-art instrument for the elemental analysis based on sample combustion and separation of gases. The N2 combustion product is quantified by a high resolution TCD detector. CO2 is trapped in the innovative Zeoquantum system while H2O is condensed by specific Permapure trap.

Fully automatically operated: from the available samplers to the oxygen dosage, from the monitoring of consumables status to the zeolite regeneration system.

ECS 8060 Model is composed by:

 ✓ Dual furnace combustion system (for a better combustion and optimization of catalysts)

 ✓ Water extraction system by Permapure and water trap

✓ Carbon dioxide adsorption system by Zeoquantum

✓ Detection system ✓ Data acquisition and handling.

Very versatile, it is very useful for analysis of a wide range of sample types and sizes - both liquids and solids, from micrograms to grams of organic materials. Thanks to the absence of a GC column, sample weight limits are wider.

Pneumatic, electric and manual samplers and high quality consumables are available. Automatisms make ECS 8060 particularly user-friendly: the automatic oxygen dosage system allows a better consumption of oxygen and consumable status monitoring function optimizes the catalysts to be used. By the Zeoquantum CO2 adsorption system a large amount of carbon dioxide can be treated by an automatic regeneration system – resulting in more precision, greater performances in a shorter time.

When used in a C-N configuration, the simultaneous determination of nitrogen and carbon is attained. The exclusive designed TCD -Thermal Conductivity Detector -is self-calibrating so that no reference gas is needed. The ECS 8060 is particularly suitable for linking to other units for determination of the isotopic ratios of carbon stable isotopes.

 

FEATURES AND BENEFITS

  • Fully automated analysis system
  • High sensitivity, accuracy and precision
  • Application flexibility and versatility
  • Zeoquantum CO2 adsorption and desorption
  • Simultaneous nitrogen and carbon determination
  • Powerful software for results visualization from PC
  • Touch-screen display for an easy settings management
  • Consumables status monitoring for an optimization of catalysts usage
  • Three types of samplers available : electronic, pneumatic, manual type
  • Easy connection to Mass Spectrometers and other detectors for carbon stable isotopes analysis
  • Low operation and management costs

 

Analytical and Technical Features

ECS 8060 Features

Type                                                      N - Proteins

Analysis Time                                        N – from 4 min up       to 8 min

Analytical Range                                  From 0,002 up to 20 mg

Accuracy*                                             <0,2% (certified standard; purity >99.9%)

Precision*                                             <0,1% (certified standard; purity >99.9%)

Sampler                                                 Pneumatic

autosampler                                                                             147 positions Electronic

autosampler                                                                             32, 50 100 positions

 Manual sampler                                                                       -

Dual Furnace System

Safety Quick-fit Integrated water trap

Zeoquantum CO2 adsorption and desorption system

 Touch-screen display

Stand-by mode

Physical Specifications

Dimensions

98x50x37cm

Weight

78kg

Power Supply

230V, 50/60Hz

Absorbed Power

5A, 1100W.

Gas Requirements

Helium (99.999% purity), 3-5 bars

 

Oxygen (99.999% purity), 3-5 bars

 

Air (oil free compressed air)

 

Analytical Conditions

 

Gas carrier

Helium

Leak test

Automatic leak test

Furnace temperature

Left furnace :1100°C max Right furnace: 1100°C max

Oxygen volume need

Automatically calculated by the oxygen doser

Flow Rate

Electronic flow rate control

Water trap

Efficient and easy maintenance

Detector

High Sensitivity TCD

Detector

High Sensitivity TCD

Detector

High Sensitivity TCD

Software data analysis

EAS Clarity

Calibrations

Linear, Quadratic, cubic

Active calibration

As necessary


Application


Accessories


Video


Catalogue and User Manual


Download Catalogue
Download User Manual