ECS 8020 - CHNS-O Analyzer Organic Elemental Analysis

ECS 8020 - CHNS-O Analyzer Organic Elemental Analysis
zoom Zoom image
Price: 0 VND
Manufacturer: NC TECH - Italy
Quantity: 

Description

ECS 8020 is a C-H-N-S-O Elemental Analyzer Model based on the Dumas combustion method.

ECS 8020 is a state-of-the-art system for the elemental analysis based on sample combustion and separation of gases with a chromatographic column.

The combustion products, i.e. CO2, H2O, N2 and SO2, are separated and quantified by a high resolution TCD detector.

ECS 8020 is composed by:

Dual furnace combustion system (for a better combustion and optimization of catalysts)

Safety quick fit system for reactors connection (easy and safe way to set the instrument)

Customized GC column

Detection system

Data acquisition and handling

ECS 80 SERIES ECS 8020 CHNS-O Several configurations may be set for the determination of the target element.

Typical configurations are for: CHNS; CHN; CN; CNS; O

Choose the best configuration by choosing the right chemicals, consumables and prepacked reactors.

The versatility of ECS 8020 can be expressed as:

Sample weight

Sample type

Liquid or solid sample

 

Analytical and Technical Features

ECS 8020 Features

 

Type Analysis time

 

CHNS-O CN

 

 

5 min

 

CHN CHNS

8 min 10/25 min

 

Analytical range

O C H

4 min

0.002-20 mg

0.002-5 mg

 

N S

0.002-20 mg

0.002-6 mg

 

Accuracy* Precision*

O

<0.2% (certified standard; purity >99.9%)

<0.1% (certified standard; purity >99.9%)

0.002-2 mg

Sampler

Pneumatic autosampler Electronic autosampler Manual sampler

147 positions max

32, 50, 100 positions

Dual furnace system Safety quick fit system Touch screen display Standby mode

 

 

 

 

 

Physical Specifications

Dimensions Weight

81x50x37 cm 68 kg

Power supply Adsorbed power

230V, 50/60Hz

5A, 1100W

Gas requirements

Helium (99.999% purity), 3-5 bar

Oxygen (99.999% purity), 3-5 bar Air (oil free compressed air)

 

Analytical Conditions

 

Gas carrier Leak test

 

Helium Automatic

Furnace temperature

Left Furnace : max 1100°C RightFurnace : max 1100°C

Oven temperature Oxygen volume need Flow rate

Max 110°C

Automatically calculated by the oxygen doser Electronic Flow Rate

Gas separation Detector

0.8-4 m GC Column High Sensitivity TCD

Software data analysis Calibration

Active calibration

EAS Clarity

Linear, Quadratic, Cubic As needed

 

Sample

 

Sample size

 

0.1-500mg (depending on sample nature)

 

Up to 1 g for soil samples

Sample type

Liquid

 

Solid

Capsule

High purity tin and/or silver capsule

 

 

Accessories

 

Installation kit Sampling kits

 

 

Small samples; big samples; deluxe kit

Microbalance Consumables

 

Technical assistance

 


Application


Accessories


Video


Catalogue and User Manual


Download Catalogue
Download User Manual